Linearity Testing
86
Applying a linearity test sample
Once the meter has detected the test strip, you will be
prompted to apply the linearity solution.
1 Wait until the ashing drop appears in the display
before applying the solution.
2 Apply a drop of the linearity solution to the front
edge of the test strip. Do not apply the solution to
the top of the strip.
The linearity solution is pulled into the test strip
area by capillary action.
Once sufcient linearity solution has been detected, the
meter beeps and the measurement begins.