Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
69
The parameters for control solutions consist of four
sepa rate values.
5 Use the keypad to enter the desired values one
after another:
Minimum limit value for Level 1
Maximum limit value for Level 1
Minimum limit value for Level 2
Maximum limit value for Level 2
6 Touch
to conrm each separate entry and
continue to the next entry.