Roche ACCU-CHEK Inform II Manual del operador Pagina 60

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 192
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 59
Glucose Control Testing
60
Results screen
The hourglass icon indicates the test is running. When the
test is completed and the result is ready, the meter beeps
again.
Depending on conguration, the result is displayed either
as a value or only as a qualitative result Pass or Fail. Your
system may be congured to disallow further testing until
all the required glucose control levels are successfully run
(QC Lockout).
When the results are displayed as a value, the results
screen contains a button that changes its name based on
the result (Range or Out of … Range). Touch this button
to display the minimum and maximum target values for
the control levels.
You can add comments to a test result (as with blood
glucose tests) (see page 50).
If you do not wish to add a comment to the test result,
touch
to continue to the next level of the control test,
if necessary, or to return to the Main Menu.
Test results are also saved when the meter is powered
off in the results screen or when it automatically
powers off.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives.
Vista de pagina 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 191 192

Comentarios a estos manuales

Sin comentarios