Initial Startup
122
Unlocking a Download Lockout
When using a data management system for congura tion,
a Download Lockout can be congured. This lockout
prevents a meter from being used for testing, if the meter
data have not been downloaded for a dened period of
time. If a download lockout occurs, the operator may use
a STAT test to temporarily override the lockout. If no STAT
test is congured or left, the meter cannot be used until
the stored data is downloaded via network (either
wire less or by docking the meter).
In cases of e.g., network failures (which make a download
impossible), or other emergencies, it may be necessary to
unlock a meter without access to the data management
system. Since unlocking is a function of the Setup Menu,
it is restricted to persons who know the setup password.
The following description assumes that the meter is
powered on and the main menu is displayed.
To unlock a meter:
1 Touch
in the Main Menu to open the Main
Menu 2 screen.
2 Touch Admin. to open the Admin Menu.
3 Touch Setup Menu to open the Setup Menu.
4 Enter the Setup password if requested.
5 Touch Unlock to reset the lockout timer.
The meter switches off, and is ready for testing after
pow ering up again.