Patient Glucose Testing
48
n The System Measurement Range refers to the
measurement range of the system itself (test strips
and meter) and is the only range which cannot be
congured. With ACCU-CHEK Inform II test strips
this xed range is 10 to 600mg/dL (0.6 to 33.3
mmol/L). If a result falls outside this range, the
message HI or LO appears, i.e., the result cannot be
quantied properly.
n Glucose results above or below the Reportable
Range, as dened by the institution or local
regulatory body, are above the highest or below the
lowest numerical results to be reported. Results
outside this range must not be used for
intervention decisions.
n Glucose results above or below the Critical Range,
as dened by the institution, require immediate
action as dened by hospital policy.
n Glucose results within the Normal Range, as
dened by the institution, are considered normal
and require no therapeutic action.