General Product Information
149
Subject/Attribute Range Default Device DMS
* Depends on available DMS settings.
Patient ID entry mode 0: Keyboard / Scan
1: List / Keyboard / Scan
2: Scan only
3: Prompt (numeric)
0 Y Y
Patient ID validation mode
(Allowed characters: a-z, 0-9,
“.” [period], “-” [hyphen])
0: none
1: length
2: list
3: list allowing entry if not
on list
4: length if numeric
0 Y Y
Patient ID maximum length (used to
validate a patient ID if patient ID validation
mode is set to “Length”
0-20 20 Y Y
Patient ID minimum length (used to
validate a patient ID if patient ID validation
mode is set to “Length”
0-20 0 Y Y
Patient name length 0-25 characters N/A N N
QC algorithm
QC algorithm 0: None (always OK)
1: Last result OK
2: Time of day (DMS only)
3: Shift (DMS only)
4: Hours
5: Strip count
6: Time of day rotate
(DMS only)
7: Shift rotate (DMS only)
8: Hours rotate (DMS
only)
9: Strip count rotate
(DMS only)
0 (Y) Y
QC hours (Number of hours between
required control measurements if QC
algorithm is set to “Hours” or “hours
rotate”)
0 – 9,999 hours 24 Y Y
QC shift length (a PASS control remains
valid for twice this amount in time after
shift start if the QC algorithm is set to
“Shift” or “Shift rotate”)
1, 2, 3, 4, 6, 8, 12, 24 hours 8 N Y
QC shift start (time of the shift start if the
QC algorithm is set to “Shift” or “Shift
rotate”)
0 – 23 hours 6 N Y
QC strip count (number of tests a PASS
control remains valid)
1 – 999 50 Y Y