Meter Setup
77
1 From the Main Menu, touch Setup to open
the meter settings
.
2 From the Setup menu, touch QC Settings.
3 From the QC Settings menu, touch STAT
Test Config.
4 Touch Enable to allow STAT Tests or
touch Disable to prohibit STAT Tests.
Your selection is now highlighted.
5 If you have enabled the option, touch
and to set the number of additional
tests.
6 Touch to save this setting, or
touch to exit this menu without saving
any changes.
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Setup Menu 09:15 am
04/19/2012
ID Setup
QC Settings
Options
Screen
QC Settings 09:15 am
04/19/2012
QC Lockout
Operator Lockout
STAT Test Config.
QC Range
STAT Test Config. 09:15 am
04/19/2012
Quantity:
Enable
Disable
3