
Control Testing and Quality Control
108
11 If performing more than one lev
el, select
the level for this measurement.
The hourglass icon shows that the test strip is
warming up. When the warming-up process is
complete, a further beep (provided the beeper
is enabled) indicates that you can now apply
the control solution.
The dropper icon flashes to indicate that the
meter is ready to perform the test and is wait-
ing for the sample to be applied.
At the same time a 180-second countdown
begins. You must apply the sample within this
time, otherwise you will receive an error
message.
QC Test 09:15 am
04/19/2012
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level 1
Level 2
QC Test 09:15 am
04/19/2012
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
QC Test 09:15 am
04/19/2012
180 SEC
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
Comentarios a estos manuales