General Product Specifications
131
11 General Product Specifications
Technical data
Sample material
Temperature range +15 °C to +32 °C (59 °F to 90 °F)
Relative humidity 10 to 85% (no condensation)
Maximum altitude 4300 m (14,000 feet)
Position Place the meter on a level, vibration-free surface or hold it so it is
roughly horizontal.
Measuring range %Q: 120 – 5
SEC: 9.6 – 96
INR: 0.8 – 8.0
Memory 2000 patient and 500 QC results with date and time
60 code chip records
Operator lists up to 5000 Operator IDs with corresponding 2nd ID
(e.g., Operator Name)
Patient lists up to 4000 Patient IDs with corresponding 2nd and 3rd
Patient IDs (e.g., name, date of birth)
Interface Infrared interface, LED/IRED Class 1
Barcode scanner Laser Class 1, according to EN 60825-1:2007
Battery operation 4 x 1.5 V batteries type AA
or special rechargeable battery pack for the CoaguChek XS Pro meter
or CoaguChek XS Plus meter
Power connection Power supply adapter: Input: 100-240 V / 50-60 Hz / 400 mA
Output: 7.5 V DC / 1.7 A
Number of tests
per set of batteries
Approx. 80 tests (approx. 60 tests per charge cycle when the recharge-
able battery pack is used)
Safety class III
Automatic power-off Programmable 1 … 60 minutes
Dimensions 231 x 97 x 43 mm
Weight 350 g (without batteries)
Sample type Capillary whole blood or non-anticoagulated venous whole blood.
Sample size At least 8 L
Interactions Refer to test strip package insert