Control Testing and Quality Control
106
7 The test strip icon prompts y
ou to insert a
test strip. Remove a test strip from its con-
tainer and close the container again with
the stopper.
8 Hold the test strip so the lettering with the
test name is facing upward.
9 Slide the test strip into the test strip guide
in the direction indicated by the arrows.
Slide the test strip in as far as it will go. A
beep tone indicates that the meter has
detected the test strip (provided the
beeper is enabled).
Exposure to external influences (such as humidity) may deteriorate the test strips and may lead
to error messages. Therefore, always close the container immediately after removing a test
strip.
QC Test 09:15 am
04/19/2012
Op: Schulz
Par: PT