Control Testing and Quality Control
105
4 Check the ba
ttery level.
If the battery icon turns red (one bar left),
there may not be enough power left for
another test.
If there are no bars left in the battery icon,
you cannot perform any more tests. Power
the meter off using the button.
In both cases, restore power by inserting
new batteries, recharging the battery
pack, or using the power adapter.
5 Check that the date and time are correct.
Correct any wrong entries as described in
Chapter 4, Meter Setup/Setting the date.
6 Touch Control Test.
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test