Meter Setup
68
QC Settings setup
The QC Settings menu contains options for enforcing the performance of liquid quality controls
by the operator at specified intervals. It also provides the option of customizing the QC Range in
accordance with applicable local guidelines.
QC Range
There are two options:
Default Range: The meter displays the QC Range provided by Roche in the code chip.
Custom Range: The option Custom Range lets the user define their own QC Range within
the default range.
Lockouts (QC Lockout and Operator Lockout)
If the liquid quality control test is not performed correctly, or if the result is outside the target
value range, the meter is locked from further use. The Lockout can also be set up selectively for
individual operators.
A liquid quality control test must be completed successfully before the meter is available again
for testing (either by the operator or in general).
The option of setting up an Operator Lockout is available only when operator lists are created
on a data management system (DMS), stored in the meter, and Operator login is activated.
These lists are only available in connection with a data management system. For more details
see Data handling, starting on
page
119.