Meter Setup
70
If y
ou selected Custom Range, the QC Range
screen opens and offers you the following
options:
■ Display target value (On/Off)
■ Deviation from target value (percentage
value). See page 69.
For the control solutions, the target value
always comes from the information stored
in the code chip. If you have chosen
Custom Range, you can now select an
allowed deviation from target value in the
range of 0 to 22.5% (in the illustration,
11.5%).
Note: Percentage deviation from target
value is always defined using INR values.
Even if you have activated % Quick in
the setup, the meter always calculates
the deviation using INR values; this
range is then converted to % Quick in a
second step. As there is no linear correla-
tion between values expressed as INR and
values expressed as % Quick, if you calcu-
late the percentage deviation (11.5 % in
the example) using the % Quick target
value, your result may deviate from the
result calculated by the meter using the
INR target value.
5 Touch On (the target value will be dis-
played with a test result) or Off (the target
value will not be displayed). Your selec-
tion is now highlighted.
QC Range 09:15 am
04/19/2012
Custom Range
Default Range
QC Range 09:15 am
04/19/2012
Deviation from
target value
Display
target value
On Off
11.5 %