Error messages
92
Error Description/Solution
Error: Time for application
exceeded
The maximum allowed time of 180 seconds for applying blood
has been exceeded.
Solution
Turn the meter off, remove the test strip and start a new
test using the same test strip and blood taken from a new
fingerstick from a different finger.
Error: Blood application
Error applying blood to the test strip.
Solution
Turn the meter off and remove the test strip. Re-read the
instructions about applying blood and repeat the test with a
new test strip and blood taken from a new fingerstick from a
different finger.
Error: Measurement
Measurement error
Solution
Turn the meter off and remove the test strip. Repeat the test
with a new test strip and blood taken from a new fingerstick
from a different finger. Do not touch or remove the test strip
when a test is in progress.
Error: Quality Control
The test strip failed the internal quality control check. The test
strip is unusable.
Solution
Turn the meter off and remove the test strip. Repeat the test
with a new test strip.
Troubleshooting error messages during or after blood application