Error messages
91
Error Description/Solution
Error: Cover
The test strip guide cover is not properly closed.
Solution
Close the test strip guide cover.
The error message disappears. The meter is ready.
Error: Test strip
The test strip is unusable or the test strip is not a CoaguChek
XS PT test strip.
Solution
Remove the test strip and repeat the test with a new or
appropriate test strip.
Error: Code chip
The code chip is missing, not properly inserted in the meter,
or damaged.
Solution
Insert the code chip, or remove it and re-insert it.
Error: Test strip
The test strip is beyond its expiration date.
Solution
First check whether the date setting is correct in the meter.
If it is not, set the correct date.
If the date is correct: Turn the meter off, remove the code
chip and test strip. Use a test strip from a new lot of test
strips and insert the code chip that came with the new lot.
Error: Test strip
The test strip is unusable.
Solution
Turn the meter off, remove the test strip and re-insert it. If the
error message appears again, discard the unusable test strip
and use a new one.
Troubleshooting error messages when preparing to test