Error messages
90
Error Description/Solution
Error: Test strip
A test strip was already inserted when the meter was turned
on with the ON/OFF button.
Error
Solution
Remove the test strip.
The error message disappears. The meter is ready.
Error: Ambient
temperature
The meter is too cold or too warm to measure correctly.
Error
Solution
Turn the meter off (unplug the power adapter) and allow it to
stand for about 30 minutes at room temperature (65°F to 90°F
or 18°C to 32°C).
Please note: When in continuous operation using the power
adapter or when recharging the battery pack, the meter can
generate the Error: Ambient temperature error message
even when the ambient temperature is under 90°F (32°C).
Error: Battery
Battery level too low
Error
Solution
Insert new batteries.
Troubleshooting error messages after turning the meter on