QC (Quality control) lockout
QC Lockout ensures that the optional liquid quality control tests are run on a regular basis for
the meter to be used. This is independent of the operator, however. Use this option if you work
without operator lists. It is also possible to mandate Operator (OP) Lockout and QC Lockout in
parallel. In addition to defining time intervals, QC Lockout can also be used to specify that a
liquid quality control test must be carried out when a new test strip lot is used (new code).
Meter setup
51
1 From the main menu screen, touch
Setup.
2 From the Setup menu screen, touch
Lockouts.
3 Touch QC Lockout.
4 Make a selection, based on lot and/
or time, for requiring a liquid quality
control test.
You may select from the following options:
n New Code Yes/No (applies every time a
new test strip lot is used)
The general intervals are:
n No
n Daily
n Weekly
n Monthly
Main Menu
09:15 am
09/16/2009
Control Test
Review Results
Setup
Patient Test
Setup Menu
09:15 am
09/16/2009
ID Setup
Lockouts
Options
Screen
Lockouts
09:15 am
09/16/2009
QC Lockout
Operator Lockout
QC Lockout
09:15 am
09/16/2009
New Code
Y e s No
Daily
Weekly
Monthly
No