8
Display Shows Possible Cause(s) Possible Solution(s)
E-54 Not Enough
Sample
j Not enough blood or control
solution was drawn into the
test strip for measurement or
was applied after the test has
started.
j Discard the test strip and repeat the test.
E-55 Code Key
Expired
j The code key is from an
expired lot of test strips.
j This message may appear when using a white
activation chip in the meter. It means the test strips
expire at the end of the current month. At the end of
the month, discard the white activation chip and
any remaining test strips. Insert a black activation
chip. Make sure the time and date in the meter are
correct. If you need a black activation chip, please
contact Roche.
E-56 Sample
Applied Early
j Blood or control solution was
applied to the test strip before
the Apply Sample screen
appeared on the display.
j Discard the test strip and repeat the test with a new
test strip.
E-57 Electronic
Error
j An electronic error has
occurred or, in rare cases, a
used test strip was removed
and reinserted.
j Turn o the meter and remove the batteries. Wait at
least 30 seconds prior to reinserting the batteries.
Turn on the meter and perform a blood glucose or
control test. If the problem persists, contact Roche.