8.2 Troubleshooting the Accu-Chek Aviva Combo Meter
Display Shows Possible Cause(s) Possible Solution(s)
The display is blank
or the meter will
not turn on.
j Batteries are dead. j Install new batteries. See Chapter 7, “Care and
Maintenance.
j Display is damaged. j Contact Roche.
j Meter is defective. j Contact Roche.
j Extreme temperaturethe
temperature is outside the
meter operating range.
j Move the meter to an area with proper temperature.
Wait ve minutes before turning on the meter. Do
not artificially heat or cool the meter.
Bluetooth icon
flashing
j Communication ended due to
a button press on the pump.
j Check the pump and continue operation of the
pump manually.
j Ensure the pump has Bluetooth wireless technology
turned on.
j Meter and pump are outside
of communication range.
j Ensure pump is within communication range.