231
8
Display Shows Possible Cause(s) Possible Solution(s)
Bolus Advice Not
Setup
j
Advice is not provided unless it
is set up through Bolus Advice
in Settings Menu.
j
You can continue to use the meter without bolus
advice or see Chapter 5, “Changing Meter Settings”
for instructions on how to set up bolus advice.
Bolus Advice
Timeout
j
Bolus advice is no longer
available for this bG result.
j
Review data that was saved in My Data. Then, decide
to modify data, add data, or begin a new bolus advice
session.
Bolus Delivery
Unavailable
j The meter cannot
communicate with the pump.
j Retry or set to Manual Pump.
j Ensure the meter and pump have Bluetooth wireless
technology turned on.
j Ensure the meter and pump are within
communication range.
j If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j The pump is currently in Stop
mode.
j Take the pump out of Stop mode.
j The pump is currently
delivering a bolus.
j Allow the current bolus delivery to complete before
delivering the next bolus.