Blood Glucose Meter Training1

Pagina 2 - Disclaimer

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter1. Make sure the meter is OFF2. Turn the meter over3. Remove the old code chi

Pagina 3 - What is changing?

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter cont’d4. Turn the code chip over so the number faces down5. Push chip into t

Pagina 4 - Learning Outcomes

Accu-Chek Performa Operator Training © VIHA 2013 Code Chip• Each box of test strips contains a code chip• Ensure this new chip is inserted i

Pagina 5

Accu-Chek Performa Operator Training © VIHA 2013 Cleaning & Disinfecting the MeterClean the meter after each client test with a Percept

Pagina 6 - • 5-second test time

Accu-Chek Performa Operator Training © VIHA 2013 Meter Cleaning Steps1. Put on clean gloves (and gown if required). 2. Make sure meter is of

Pagina 7 - Arrow Buttons

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strips15

Pagina 8 - Changing the Battery

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strip - Features• 0.6 µL sample size• Wipe 1stdrop and use 2nddrop•

Pagina 9 - Changing the Battery cont’d

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Control Solution• Only use Accu-Chek Performa Control Solutions• Write t

Pagina 10 - Coding a meter

Accu-Chek Performa Operator Training © VIHA 2013 Why do a Quality Control ?• Performing a Quality Control verifies the reliability of the te

Pagina 11 - Coding a meter cont’d

Accu-Chek Performa Operator Training © VIHA 2013 When to do Quality ControlA Quality Control should be done:• Immediately prior to client te

Pagina 12 - Code Chip

Accu-Chek Performa Operator Training © VIHA 2013 Disclaimer• This information is intended for use by Health Care Professionals • Patients wh

Pagina 13 - Remember:

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – for Healthcare Professionals 1. Insert the test strip into the

Pagina 14 - Meter Cleaning Steps

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d3. Select the control solution to be used4. Place the met

Pagina 15

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d6. Touch the second drop to the front yellow edge of the

Pagina 16 - • Wipe 1

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d– The QC result appears on the display, along with a cont

Pagina 17

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d10. Result is in range if “OK” & the result alternate

Pagina 18 - Why do a Quality Control ?

Accu-Chek Performa Operator Training © VIHA 2013 Troubleshooting - Quality Control is Out of RangeReview this list for possible causes:– Wer

Pagina 19 - When to do Quality Control

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test 1. Insert a test strip, the meter turns on 2. Make sure the

Pagina 20 - Professionals

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d3. A test strip and flashing blood drop appear on t

Pagina 21 - 4. Place the meter on

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d5. Wipe the first drop away and touch the second dr

Pagina 22 - (means that enough sample

Accu-Chek Performa Operator Training © VIHA 2013 Hi and Low Results Display• Hi Level - greater than 33.3 mmol/L• Low Level – less than 0.6

Pagina 23

Accu-Chek Performa Operator Training © VIHA 2013 What is changing?Roche ACCU-CHEK® meters will be the meters used by all Health Authorities

Pagina 24

Accu-Chek Performa Operator Training © VIHA 2013 Reference RangesVIHA Glucose Meter Normal Reference Interval 4.0 - 11.0 mmol/L(The meter wi

Pagina 25

Accu-Chek Performa Operator Training © VIHA 2013 VIHA Critical Values(follow your facilities procedures for appropriate actions)Critical Hig

Pagina 26 - 2. Make sure the code number

Accu-Chek Performa Operator Training © VIHA 2013 Documenting ResultsRemember - Always document the result as soon as possible to avoid error

Pagina 27

Accu-Chek Performa Operator Training © VIHA 2013 Viewing Results - AveragesThe Perform meter has the ability to show 7, 14 or 30 day averag

Pagina 28

Accu-Chek Performa Operator Training © VIHA 2013 Limitations of ACCU-CHEK®Test Strips (Refer to test strip package insert for further inform

Pagina 29 - Hi and Low Results Display

Accu-Chek Performa Operator Training © VIHA 2013 Maintaining Competency Competency on the blood glucose meter is your professional responsi

Pagina 30 - Reference Ranges

Accu-Chek Performa Operator Training © VIHA 2013 Learning OutcomesThese slides should help you understand: • Basic meter features • How to p

Pagina 31 - Critical High Critical Low

Accu-Chek Performa Operator Training © VIHA 2013 Review of existing BGM practice/policy Client identification must be confirmed using a min

Pagina 32 - Documenting Results

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter Features• 5-second test time• 0.6 – 33.3 mmol result range• Large

Pagina 33 - Viewing Results - Averages

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter FeaturesDisplayRight and Left Arrow ButtonsOn/Off/Set ButtonBatter

Pagina 34 - Key Points:

Accu-Chek Performa Operator Training © VIHA 2013 Changing the BatterySteps:1. Open the battery door on the back of the meter by pushing the

Pagina 35 - Maintaining Competency

Accu-Chek Performa Operator Training © VIHA 2013 Changing the Battery cont’d93. Insert the new battery with the + side up.4. Put the battery