Inform II test strip
Technical specications
Accu-Chek Inform II system components
Specication Test strip
Measurement principle Modied glucose dehydrogenase (Mut. Q-GDH) - electrochemical
detection technique
Range of measurement 10 to 600 mg/dL (0.56 to 33.31 mmol/L)
Measurement time 5 seconds
Operating temperature 8 to 44 °C (46 to 111 °F)
Sample volume 0.6 µL
Hematocrit range 10 to 65%
Altitude < 3,094 meters above sea level (10,150 feet)
Sample types Capillary, venous, neonatal, arterial
Calibration IFCC plasma
Lancets
“Pain free” single-use
lancing device offering
three adjustable depth
settings
Test strips
0.6 μL sample size
5 second test time
End dosing for
easier testing
Test strip vial
Flip top vial to
maintain integrity
of test strips
18 months open
vial stability
Control and
linearity solutions
Ensures function
and accuracy
2 levels of controls/
6 levels of linearity
Code key reader
Entry of strip information
into device. Ensures only
approved test strips are
used for testing
Accessory box
Convenient storage
and transport of meter
and consumables
Base Unit
Recharges meter battery.
Allows for the transfer of data
to and from the data man-
agement system
cobas IT 1000 application