Display Shows Possible Cause(s) Possible Solution(s)
Bolus Delivery Unavailable j The meter cannot communicate
with the pump.
j Retry or set to Manual Pump.
j Ensure the meter and pump have
Bluetooth wireless technology turned on.
j Ensure the meter and pump are within
communication range.
j If the low battery icon is displayed on
the meter, the meter cannot
communicate with the pump. Replace
the batteries.
j The pump is currently in Stop
mode.
j Take the pump out of Stop mode.
j The pump is currently delivering a
bolus.
j Allow the current bolus delivery to
complete before delivering the next
bolus.