Roche Accu-Chel Manual de servicio Pagina 75

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 135
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 74
75
Checking your meter
7 Open a bottle of control solution.
8 Wipe the tip of the dropper with a clean, dry
paper towel to remove any dried-on residues.
9 Invert the bottle and hold it downwards at an
angle. Squeeze it gently until a small drop which
is free of any air bubbles is suspended from the
tip.
10 Apply one drop to the centre of the orange eld
without touching it with the tip of the dropper.
The eld must be completely covered with
solution.