Roche Accu-Chek Compact Plus Manual de usuario Pagina 82

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 94
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 81
160
Farbe: P 249 – Prüfmittelnummer: 126 X
Symbols, display messages and troubleshooting
161
Symbols, display messages and troubleshooting
10
This message/symbol
appears…
and means:
A meter error has occurred.
Press the Test button to turn the me-
ter off and then on again. If the error
message persists, please contact
your local customer support and ser-
vice centre (see Chapter 16 “Local
customer support and service”).
During a data download both arrows fail
to start flashing. The meter turns itself
off after 5 minutes and the PC, hand
-
held computer or the Accu-Chek Smart
Printer System also displays an error
message.
A download error occurred.
Start the download again.
A strong light source is too close to
the two infrared windows.
Move away from the light source
and start the download again.

This message/symbol
appears…
and means:

The infrared windows are too far
apart or not properly facing one an
-
other.
Reduce the distance between the
two infrared windows to 10-20 cm
and position them so they are fac-
ing one another. Start the download
again.
A software error occurred on the PC,
handheld computer or the
Accu-Chek Smart Printer System.
Read the manual for the software or
the Accu-Chek Smart Printer Sys
-
tem.
Vista de pagina 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 93 94

Comentarios a estos manuales

Sin comentarios